Place of Origin: | China |
Brand Name: | FLUKE |
Model Number: | 743B |
Minimum Order Quantity: | 1 |
---|---|
Price: | negotiable |
Packaging Details: | Standard FLUKE packing with larger carton to protect the materials |
Delivery Time: | 6-8 weeks |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 500pcs per week |
Warranty Time: | 12 Months | Lead Time: | 8-10 Weeks |
---|---|---|---|
Packing: | Standard FLUKE Packing With Larger Carton To Protect The Materials | Quantity: | 100% Original And Genuine |
Place Of Origin: | China |
Fluke 743B Documenting Process Calibrator with Data Logging
overview
The 743B does the work of several tools – sourcing, simulating and measuring pressure, temperature, and electrical signals in one rugged, hand-held device.
For documentation, the 743B automates calibration procedures and captures your data.
And, of course, it helps you meet rigorous standards like ISO 9000, FDA, EPA and OSHA regulations.
The 743B has a PC interface that lets you download procedures, lists, and instructions created with software-or upload data for printing, archiving, and analysis. With its expanded memory, the 743B can hold a full week of calibrations and procedures.
Fluke 743B Documenting Process Calibrator with Data Logging
The 743B is easy to use with flexible test strategies. And like all Fluke tools, it's rugged, reliable and tough as nails. It's like having several powerful tools in one - so you can work faster than ever:
Fluke 743B Documenting Process Calibrator with Data Logging
Measurement Accuracy | |||
Voltage DC | 110.000 mV | 0.025% + 0.015%1 | |
1.10000 V | 0.025% + 0.005%1 | ||
11.0000 V | 0.025% + 0.005%1 | ||
110.000 V | 0.05% + 0.005%1 | ||
300.00 V | 0.05% + 0.005%1 | ||
Voltage AC | 20 to 40 Hz | 2% + 10 | |
40 to 500 Hz | 0.5% + 5 | ||
500 to 1 kHz | 2% + 10 | ||
1 kHz to 5 kHz | 10% + 20 | ||
Ranges | 1.1000, 11.000, 110.00, 300 V | ||
Current DC | 30.000 mA | 0.01% + 0.015%1 | |
110.00 mA | 0.01% + 0.015%1 | ||
Resistance | 11.000 Ω | 0.05% + 50 mΩ | |
110.00 Ω | 0.05% + 50 mΩ | ||
1.1000 kΩ | 0.05% + 500 mΩ | ||
11.000 kΩ | 0.1% + 10 Ω | ||
Frequency | 1.00 to 109.99 Hz | 0.05 Hz | |
110.0 to 1099.9 Hz | 0.5 Hz | ||
1.100 to 10.999 kHz | 5 Hz | ||
11.00 to 50.00 kHz | 50 Hz | ||
Pressure | Accuracy | From 0.025% of range using any of 29 pressure modules. | |
(For detailed specifications refer to pressure modules in options and accessories) | |||
Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |||
1. (% of reading + % of full scale) | |||
Source Accuracy | |||
Voltage DC | 110.000 mV | 0.01% + 0.005%1 | |
1.10000 V | 0.01% + 0.005%1 | ||
15.0000 V | 0.01% + 0.005%1 | ||
Current DC | 22.000 mA (Source) | 0.01% + 0.015%1 | |
22.000 mA (Simulate) | 0.02% + 0.03%1 | ||
Resistance | 11.000 Ω | 0.01% + 20 mΩ | |
110.00 Ω | 0.01% + 40 mΩ | ||
1.1000 kΩ | 0.02% + 500 mΩ | ||
11.000 kΩ | 0.03% + 5 Ω | ||
Frequency | 0.00 to 10.99 Hz | 0.01 Hz | |
11.00 to 109.99 Hz | 0.1 Hz | ||
110.0 to 1099.9 Hz | 0.1 Hz | ||
1100 to 21999 Hz | 2 Hz | ||
22.000 to 50.000 kHz | 5 Hz | ||
1. (% of reading + % of full scale) | |||
RTDs and Thermocouples | |||
Measure accuracy | 10 Ω Cu (427) | 2°C | |
100 Ω Pt (3916) | 0.3°C | ||
100 Ω Pt (3926) | 0.3°C | ||
100 Ω Pt (385) | 0.3°C | ||
200 Ω Pt (385) | 0.3°C | ||
500 Ω Pt (385) | 0.3°C | ||
1000 Ω Pt (385) | 0.3°C | ||
120 Ω Ni (672) | 0.3°C | ||
For 2-and 3-wire measurement, add 0.4°C | |||
E | 0.3°C | ||
N | 0.5°C | ||
J | 0.3°C | ||
L | 0.3°C | ||
K | 0.3°C | ||
T | 0.3°C | ||
U | 0.3°C | ||
B | 0.9°C | ||
R | 1.0°C | ||
S | 0.9°C | ||
C | 0.6°C | ||
Accuracy with external cold junction, for internal junction add 0.2°C | |||
Source accuracy | 10 Ω Cu (427) | 1°C | |
100 Ω Pt (3916) | 0.1°C | ||
100 Ω Pt (3926) | 0.1°C | ||
100 Ω Pt (385) | 0.1°C | ||
200 Ω Pt (385) | 0.1°C | ||
500 Ω Pt (385) | 0.1°C | ||
1000 Ω Pt (385) | 0.1°C | ||
120 Ω Ni (672) | 0.1°C | ||
For 2-and 3-wire simulation, add 0.4°C | |||
E | 0.2°C | ||
N | 0.3°C | ||
J | 0.2°C | ||
L | 0.2°C | ||
K | 0.3°C | ||
T | 0.3°C | ||
U | 0.3°C | ||
B | 0.8°C | ||
R | 0.9°C | ||
S | 0.9°C | ||
C | 0.6°C | ||
Accuracy with external cold junction, for internal junction add 0.2°C | |||
Specifications | |||
Data log functions | Measure functions | Voltage, current, resistance, frequency, temperature, pressure | |
Reading rate | 1, 2, 5, 10, 20, 30, or 60 readings/minute | ||
Maximum record length | 8000 readings (7980 for 30 or 60 readings/minute) | ||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature | |
Rate | 4 steps/second | ||
Trip detect | Continuity or voltage (continuity detection not available when sourcing current) | ||
Loop power function | Voltage | Selectable, 24 V or 28 V | |
Accuracy | 5% | ||
Maximum current | 22 mA, short circuit protected | ||
Maximum input voltage | 30 V DC | ||
Step functions | Source functions | Voltage, current, resistance, frequency, temperature | |
Manual step | Selectable step, change with arrow buttons | ||
Autostep | Fully programmable for function, start delay, stepvalue, time per step, repeat | ||
Environmental Specifications | |||
Operating temperature | -10°C to +50°C | ||
-20°C to +50°C2 | |||
Storage temperature | -20°C to +60°C | ||
Dust/water resistance | Meets IP52, IEC 529 | ||
Operating altitude | 2800 m | ||
2. Except frequency and AC | |||
Safety Specifications | |||
Agency approvals | CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-11993 | ||
Mechanical and General Specifications | |||
Size | 130 x 236 x 61 mm | ||
Weight | 1.4 kg | ||
Batteries | NiCd 7.2V, 1.7 Ah | ||
Battery life | ~8 hours typical | ||
Battery replacement | Via snap-shut door without opening calibrator; no tools required | ||
Side port connections | Pressure module connector | ||
RS-232 connector for PC interface cable | |||
Connection for optional battery eliminator | |||
Data storage capacity | 1 week of calibration results | ||
90 day specifications | The standard specification intervals for the 743B is 1 and 2 years. | ||
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. Floor specifications, expressed as "% of full scale" or"counts" or "ohms" remain constant. |
contact
Why do you choose us?
1. 100% original package product, 100% quality assurance.
If you find a fake product. Please contact us without hesitation! We afford the shipment cost! We send you a new product asking zero cent!
2. Varied payment methods.
We adopt Paypal, T/T, Alibaba Trade Assurance order, and other methods. If you can not trust us , you can choose to pay it the other way! 100% make your money safe!
We mainly supply Yokogawa, Emerson, Rosemount, Honeywell, Siemens, ABB, FLUKE,Endress+Hauser,SchneiderElectric,Moeller,FUJI,Omron,+,Hima, Yaskawa,Norgren, CHINT,Hach,Weidmuller,Phoenix,MTL,SMC,Testo,Rockwell ect.
Wuhan Xieyuan Automation Equipment Co., Ltd
Telephone:86-027-85583233
Mobile Phone:86-18671080257
Fax:86-027-85309780
Contact : victor@whxyauto.com
Skype:victordu963@gmail.com
Whatsapp & Viber are available
Address:7-A16,Caishen Commercial Plaza,Hankou Railway Station